Libra 40W 14” X 17” Wireless

Wireless, Cassette Size, Flat Panel Detector

Excellent Sensitivity & Resolution

・Advanced and proven fine structured CsI:Tl and direct vapor deposition technology deliver higher sensitivity and resolution.

・Reflection coating on CsI:Tl screen enables excellent Detective Quantum Efficiency (DQE) and high Modulation Transfer Function (MTF).

・Lower radiation dose beneficial to patients as a result of excellent image quality. The Libra 40W offers a new level of functionality and reliability for system manufacturers.

Excellent Reliability

・Excellent durability by using CsI:Tl screen direct vapor deposition method.

・The structure is highly reliable and protected from degradation due to the use of a unique moisture-proof sealing method for the CsI:Tl screen. 

Specifications

Specifications

Sensor Unit
Sensor Protection Plate Carbon Fiber Plate
Cooling Natural Air Cooling
Input DC16V (from Power Supply)
Power Consumption

Maximum 13W

Overall Dimensions 383.5mm×459.5mm×15mm (W(H)×D(V)×(H))
Weight 3.1 kg (approx.)
Power Supply: DEPS9601
Input AC100 to 240V,1.0A,50/60Hz 
Output DC16V 3A
Overall Dimensions 155mm×208mm×65mm (W(H)×D(V)×(H)) 
Weight 1.2 kg (approx.)
Main Cable: MI39-01545A
Length 7m
Diameter φ8~9mm
Li-ion Battery: SDB-3S1PA
Nominal Capacity 3400 mAh
Nominal Voltage 11.45V
Outline Dimensions 245mm×130mm×8.5mm (W(H)×D(V)×(H))
Weight 385g
Battery Recharger: DEPS-9606
Outline Dimensions 281.6×206.6×21.2 mm (W(H)×D(V)×(H))
Weight 500g (approx.)
AC Adapter:EPS-F007A(Model No.: SPU100-107)
Outline Dimensions 76×146×43 mm (W(H)×D(V)×(H))
Cable Length 430mm
Input AC100 to 240V 1.2-0.5A 47-63Hz
Output DC19V 5.26A
Weight

510g (approx.) 

Image Format
X-ray Conversion Layer Cesium Iodide (Csl) with Amorphous Silicon (a-Si) Photodiode
Active Area 345(H)×423(V)mm (14×17 inch)
Pixel Matrix 2466(H)×3040(V) 
Pixel Pitch 140μ m
Cycle Time

Shot to Shot 12 sec (WLAN)

Shot to Shot 9 sec (Ethernet)

(Cycle time is the time to complete image transfer from the X-ray Exposure. Cycle time does not include image processing time. The image processing time is determined by the specifications of the image processing unit.) 
Performance
Limiting Resolution 3.7 Lp/mm typ.
MTF (2.0Lp/mm, 70kVp, 1×1) 36 % typ.
DQE (DQE (0), Quantum - Limited) > 70 %
A/D Conversion 14 bit
Ratings
Energy Range 40 to 150kVp
Maximum Entrance Dose (Linear Output Range) 4mR
Interface
    Sensor Unit
        Unit Interface Connect to power supply
    Power Supply
        Unit Interface Connect to Sensor Unit
        Data Output 14 bit Digital Output Ethernet (1000BASE-T)
        Command Control Ethernet (1000BASE-T)
        Power Input AC100 to 240V, 1A 50/60 Hz
    WLAN

IEEE802.11a/b/g/n

2.4GHz/5GHz 

Image Acquisition Exposure period
X-ray period (ms) Standard: 500 (500,1000,1500,2000, 2500, 3000, 3500, 4000)