PRINT CARDS

1.800.817.5030

Catalog

Krystal Rad 1000

Digital Radiographic System Featuring a Height Adjustable Rotating Swivel Arm and Compact Floor to Wall Mount Design.

Category:

Fully Motorized Universal Radiographic System

The Krystal Rad 1000 features a sturdy ¬oor-to-wall mounting system, with a dual-speed motorized swivel arm for operational ¬exibility. Precise positioning control for all movements is accessed, controlled and monitored through the use of a 6” color touch screen and 12 key wireless remote control pad. This system enables exceptionally accurate control of source-to-image-distance, the height of the swivel C-arm, and the angulation of the swivel arm and detector. Asa patient safety precaution, all operational movements
stop when the switch is released.

Intuitive Design
The Krystal Rad 1000 Single-button operation allows for 1m and 1.8m chest positioning, or 1m table positioning, as well as easy preset angle positioning via the 6” color touch screen.
Seven of the most common radiology exams are quickly accessible via the 6” touch-screen color control monitor, allowing convenient one-touch positioning after program selection. Three laser beams and pressure sensors work in combination to provide a unique, anti-collision
system for added patient security and equipment protection.

Choice in Flat Panel Detectors
The Krystal Rad 1000 o ers several choices from out Libraseries of Cesium digital ¬at panel detectors.
– 17″ x 17″
– Wireless, tethered, and xed
– 14 and 16 bit depth
– 100μ to 143μ
All Libra ¬at panel detectors feature Automatic Exposure Detection (AED), while the Libra HD technology o ers greatly improved image quality coupled with low patient dosages.

Stand

• Horizontal Column Height – 224.6 cm
• Vertical Carriage Movement – 44.0 cm to 163.5 cm to floor with motorized movement
• Swivel Arm Rotation – 0 -90 degrees, motorized movement
• Bucky Tilt – +/- 30 degrees, mechanical lock, and 0-degree detent
• SID Movement – motorized from 1 meter to 2 meter
• Maximum Tube Height @ 0 degree rotation – 285.0 cm

Generator

• High frequency of 32 kW (maximum 450 MHz)
• kV range – 40 – 125 kV
• mA range – 10 – 400 mA
• mAs range – 0.1 – 500 mAs
• Technique Selection – kV/AEC, kV/mAs, kV/mA/ms

DR Detector

• 17 x 17 Csi Fixed Panel
• Image format – 9.2M Pixels (3008 x 3072)
• Pixel Pitch – 143 micron
• A/D Conversion – 14 bit

X-ray Tube

• Nominal Radiographic tube Voltage – 40 -125 kV
• Nominal Focal Spot – Large 1.2 mm, Small 0.6 mm
• Nominal Anode heat Input Energy – Large 30 kW, Small 15 kW
• Anode target Angle -12 degrees
• Anode Heat Content -140 kHU

Workstation and Monitor

• 3rd Gen Intel Core
• Windows 7 Professional, 64-bit
• 4GB 2DDR3 SDRAM at 1600 MHz
• 500GB 3.5 SATA 6Gb/s with 16 MB DataBurst Cache
• Monitor – 21.5” (54.6cm) diagonal 10

Top